GENERAL PERFORMANCE
- Model: fully automated, discrete, random access, bench-top, STAT sample priority
- Throughput:
- 240 T/H constant speed
- Up to 400 T/H with ISE
- Measuring principles: colorimetry, turbidimetry, ISE method
- Methodology:
- Endpoint, fixed-time, kinetic, ISE optional
- Single/dual reagent
- Monochromatic/bi-chromatic
- Linear/non-linear multipoint calibration
- Reagents: kits ready to use
- Programming: closed system with user defined profiles and calculation parameters
- Auto QC, auto calibration, reflex test
- Sample tray: 50 positions (option to extend up to 100 positions) in 24-hour
- refrigerated compartment
- Reagent tray: 50 positions in 24-hour refrigerated compartment
- Sample volume: 2-35 μl, step by 0.1 μl
- Reagent volume: 10-200 μL, step by 0.5 μL
- Reagent/Sample probe: liquid level detection, horizontal and vertical collision
- protection, inventory checking, reagent pre-warming, clog detection
- Probe cleaning:
- Automatic washing for interior and exterior
- Probe carry-over ≤ 0.05%
- Automatic sample dilution: pre-dilution and post-dilution
- Barcode reader:
- Built-in version
- Used for sample and reagent programming
- Capable to communicate with LIS in bi-directional mode
- Reaction rotor: rotating tray, containing 80 cuvettes
- Cuvette: reusable, optical length 5 mm
- Reaction volume: 100-360 μl
- Cuvette washing: washing station with pre-warmed detergent and water
- Operating temperature: 37°C
- Mixing unit: independent mixing bar
- Water consumption: ≤ 6.5 L/H
- Light source: halogen-tungsten lamp
- Absorption range: 0~3.5 Abs
- Wavelength: grating photometer with 12 wavelengths (340 nm, 380 nm, 412 nm, 450nm, 505 nm, 546 nm, 570 nm, 605 nm, 660 nm, 700 nm, 740 nm, 800 nm)
- Optional, measuring: K+ , Na+, ClCONTROL AND CALIBRATION
- Calibration modes:
- K factor
- Linear: two points and multi-points
- Non-linear: Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial,
- Parabola, Logit-Log 3P, Broken line
- Control rules: Westgard multi rules, Cumulative sum check, Twin-plot, Levey-Jennings
- Software: Windows 10 (64 bit)
- Interface: RS-232
- Power supply: 200V-240V, 50/60Hz, ≤1300VA or 110V/115V, 60Hz, ≤1300VA
- Temperature: 15-30°C for operation, 0-40°C for storage
- Humidity: 35% -85% RH, without condensation
- Dimension: ≤ 860 mm (length) × 660 mm (depth) × 550 mm (height)
- Weight: ≤ 130 kg
Reviews